Cryo-electron microscopy structure of a bovine CLC-K chloride channel, main (class 1) conformation
Sample A bovine CLC-K channel complexed with a monoclonal antibody fragment (Fab)
Specimen Preparation Sample Aggregation State PARTICLE Vitrification Instrument FEI VITROBOT MARK IV Cryogen Name ETHANE Sample Vitrification Details
3D Reconstruction Reconstruction Method SINGLE PARTICLE Number of Particles 82167 Reported Resolution (Å) 3.76 Resolution Method FSC 0.143 CUT-OFF Other Details the indicated resolution was estimated on the primary map using the provided mask including Fab. Refinement Type Symmetry Type POINT Point Symmetry C2
Map-Model Fitting and Refinement Id 1 Refinement Space Refinement Protocol Refinement Target Overall B Value Fitting Procedure Details
Data Acquisition Detector Type GATAN K2 SUMMIT (4k x 4k) Electron Dose (electrons/Å**2) 1.8
Imaging Experiment 1 Date of Experiment Temperature (Kelvin) Microscope Model FEI TITAN KRIOS Minimum Defocus (nm) Maximum Defocus (nm) Minimum Tilt Angle (degrees) Maximum Tilt Angle (degrees) Nominal CS Imaging Mode BRIGHT FIELD Specimen Holder Model Nominal Magnification Calibrated Magnification Source FIELD EMISSION GUN Acceleration Voltage (kV) 300 Imaging Details
EM Software Task Software Package Version PARTICLE SELECTION RELION 1.4 CTF CORRECTION CTFFIND 4 MODEL FITTING Coot 0.8.1 MODEL REFINEMENT REFMAC 5.8 INITIAL EULER ASSIGNMENT RELION 1.4 FINAL EULER ASSIGNMENT RELION 1.4 CLASSIFICATION RELION 1.4 RECONSTRUCTION RELION 1.4
Image Processing CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details PHASE FLIPPING AND AMPLITUDE CORRECTION