6RKW
CryoEM structure of the complete E. coli DNA Gyrase complex bound to a 130 bp DNA duplex
ELECTRON MICROSCOPY
Sample |
---|
The complete E. coli DNA Gyrase complex bound to 130bp DNA duplex |
Sample Components |
DNA gyrase subunit A |
DNA gyrase subunit B |
DNA |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | |
Cryogen Name | ETHANE |
Sample Vitrification Details |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 94633 |
Reported Resolution (Å) | 6.6 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (3NUH, 3NUH, 3NUH, 3NUH, 1EI1, 1EI1, 1ZI0) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | |||||
Refinement Target | |||||
Overall B Value | |||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 50 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 300 |
Maximum Defocus (nm) | 800 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 0.01 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | |
Nominal Magnification | 130000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
IMAGE ACQUISITION | SerialEM | |
CTF CORRECTION | RELION | 2.0.3 |
CTF CORRECTION | Gctf | 1.06 |
MODEL FITTING | UCSF Chimera | 1.11.2 |
INITIAL EULER ASSIGNMENT | cryoSPARC | V1 |
FINAL EULER ASSIGNMENT | RELION | 2.0.3 |
RECONSTRUCTION | RELION | 2.0.3 |
MODEL REFINEMENT | PHENIX | 1.12 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION |