ELECTRON MICROSCOPY
Refinement
RMS Deviations | |
---|---|
Key | Refinement Restraint Deviation |
r_lrange_it | 264.032 |
r_mcangle_it | 136.303 |
r_scangle_it | 135.451 |
r_scbond_it | 86.101 |
r_mcbond_it | 78.101 |
r_dihedral_angle_2_deg | 33.74 |
r_dihedral_angle_4_deg | 17.49 |
r_dihedral_angle_3_deg | 16.618 |
r_dihedral_angle_1_deg | 9.059 |
r_angle_refined_deg | 1.977 |
Sample |
---|
human VPS34 complex II (VPS34-VPS15-Beclin1-UVRAG) bound to human Rab5a |
Sample Components |
human VPS34 complex II (VPS34-VPS15-Beclin1-UVRAG) |
Ras-related protein Rab-5A |
Specimen Preparation | |
---|---|
Sample Aggregation State | 3D ARRAY |
Vitrification Instrument | FEI VITROBOT MARK I |
Cryogen Name | ETHANE |
Sample Vitrification Details | Blot force 20, blot time 6 s |
3D Reconstruction | |
---|---|
Reconstruction Method | SUBTOMOGRAM AVERAGING |
Number of Particles | 26979 |
Reported Resolution (Å) | 9.8 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (5DFZ, 3MJH, 4DDP, 3IHY) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | FLEXIBLE FIT | ||||
Refinement Target | correlation | ||||
Overall B Value | 320 | ||||
Fitting Procedure | |||||
Details | An initial model was build with SWISSMODEL. using 5DFZ, 4DDP and 3IHY. This was then fit manually to density while running restrained MD with ISOLDE. ... |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K3 BIOQUANTUM (6k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 2.99 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 2000 |
Maximum Defocus (nm) | 5000 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 81000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
VOLUME SELECTION | MATLAB | R2016b |
IMAGE ACQUISITION | SerialEM | 3.8 |
CTF CORRECTION | NOVACTF | 1.0.0 |
MODEL FITTING | Coot | 0.9 |
MODEL FITTING | ISOLDE | 1.1 |
FINAL EULER ASSIGNMENT | MATLAB | R2016b |
CLASSIFICATION | MATLAB | R2016b |
RECONSTRUCTION | RELION | 3.1 |
MODEL REFINEMENT | REFMAC | 5.8.0272 |
MODEL REFINEMENT | ISOLDE | 1.1 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION |